A noise reduction scheme on digitized mammographic phantom images is presented. This algorithm is based on a direct contrast modification method using an optimal function which is obtained by means of the mean squared error as a criterion. Computer simulated images containing objects similar to those observed in the phantom are built to evaluate the performance of the algorithm. Noise reduction results obtained on both simulated and real phantom images show that the developed method may be considered as a good pre-processing step from the point of view of automating phantom film evaluation by means of image processing.
English
Noise reduction; Signal-dependent noise; Mammographic phantom; Contrast modification; Reducció de soroll; Soroll de senyal dependent; Modificació de contrast; Reducción de ruido; Ruido de señal dependiente; Modificación de contraste
ELCVIA. Electronic letters on computer vision and image analysis ; V. 5 n. 4 (2006) p. 64-74
open access
Aquest document està subjecte a una llicència d'ús Creative Commons. Es permet la reproducció total o parcial, la distribució, i la comunicació pública de l'obra, sempre que no sigui amb finalitats comercials, i sempre que es reconegui l'autoria de l'obra original. No es permet la creació d'obres derivades.
https://creativecommons.org/licenses/by-nc-nd/3.0/