dc.contributor
Universitat Politècnica de Catalunya. Departament d'Enginyeria Electrònica
dc.contributor
Universitat Politècnica de Catalunya. GRUP ISI - Grup d'Instrumentació, Sensors i Interfícies
dc.contributor.author
Pallàs-Areny, Ramon
dc.date.issued
2019-12-13
dc.identifier
Pallàs-Areny, R. Comments on “A novel high input impedance front-end for capacitive biopotential measurement”. "Medical & biological engineering & computing", 13 Desembre 2019, vol. 58, p. 267-269.
dc.identifier
https://hdl.handle.net/2117/179818
dc.identifier
10.1007/s11517-019-02083-x
dc.description.abstract
A front-end for biopotential sensing in wearable medical devices has been recently proposed which is claimed to provide 100 GOinput impedance by manually matching two resistor pairs in a positive- and a negative-feedback loop around an operationalamplifier (op amp); the cost being that the equivalent input noise voltage doubles with respect to a simple non-inverting amplifier.The ECG acquired with capacitive (sic) electrodes through a cotton shirt is presented as a proof of the performance of theproposed circuit. It turns out, however, that the analysis ignores op amp’s input capacitance hence the effort to achieve a very highinput resistance seems futile. Further, cotton is highly hygroscopic hence not an appropriate dielectric, so that there is no proofthat the electrodes tested were actually capacitive. This comment addresses these two problems and some additional conceptualand methodological inaccuracies found in the paper.
dc.description.abstract
Peer Reviewed
dc.description.abstract
Postprint (published version)
dc.format
application/pdf
dc.relation
https://link.springer.com/article/10.1007%2Fs11517-019-02083-x
dc.rights
http://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.rights
Attribution-NonCommercial-NoDerivs 3.0 Spain
dc.subject
Àrees temàtiques de la UPC::Enginyeria electrònica
dc.subject
Supercapacitors
dc.subject
Electronic circuits--Noise
dc.subject
Biopotential sensing
dc.subject
Capacitive electrodes
dc.subject
Voltage-loading effect
dc.subject
Noise analysis
dc.subject
Circuits electrònics -- Soroll
dc.title
Comments on “A novel high input impedance front-end for capacitive biopotential measurement”