Efficacy and Safety of Open-Conjunctiva Ab Externo 63 µm vs. 45 µm XEN® Gel Stent in Glaucoma Surgery: One-Year Follow-Up

Altres autors/es

Institut Català de la Salut

[Bertolani Y, Rigo-Quera J, Sánchez-Vela L, Pujol-Carreras O, Amilburu M, Dou A, Castany M] Servei d’Oftalmologia, Vall d’Hebron Hospital Universitari, Barcelona, Spain

Vall d'Hebron Barcelona Hospital Campus

Data de publicació

2025-09-29T12:28:16Z

2025-09-29T12:28:16Z

2025-05



Resum

Intraocular surgery; Minimally invasive glaucoma surgery; Open-conjunctiva approach


Cirugía intraocular; Cirugía de glaucoma mínimamente invasiva; Abordaje conjuntival abierto


Cirurgia intraocular; Cirurgia de glaucoma mínimament invasiva; Abordatge conjuntival obert


Background: To compare the efficacy and safety of the XEN® 63 µm and 45 µm devices with the ab externo open conjunctiva with a 30G needle approach. Methods: A retrospective, non-randomized and single-center study was conducted. Consecutive eyes undergoing a XEN® 63 µm implant were compared with a matched cohort of cases with a XEN® 45 µm implant. Standalone and combined procedures with phacoemulsification were included. Results: A total of 28 XEN® 45 µm and 28 XEN® 63 µm were included. Complete surgical success was achieved in 17 cases (60.7%) in the 45 µm group and in 20 cases (71.4%) in the 63 µm group, with no statistical differences. One year after the surgery, the mean IOP was 13.8 ± 3.3 mmHg for the 45 µm group and 12.4 ± 4.2 mmHg for the 63 µm group (p-value > 0.05). Likewise, the use of glaucoma medication was lowered in the 63 µm device (0.32 ± 0.87) compared to the 45 µm device (0.39 ± 0.86), with no statistical significance. Postoperative hypotony was more frequent in the 63 µm device (39.3%) than in the 45 µm group (28.6%), with no statistical differences. However, hypotony-associated complications (including choroidal detachment, hypotony keratopathy, and hypotony maculopathy) were significantly higher in the 63 µm group (p = 0.011). Conclusions: Although the XEN® 63 µm may offer a greater IOP-lowering effect with better complete surgical success, no significant differences were detected compared to the 45 µm device. Hypotony-related complications were higher in the XEN 63 µm, although most of them resolved with conservative management.

Tipus de document

Article


Versió publicada

Llengua

Anglès

Publicat per

MDPI

Documents relacionats

Journal of Clinical Medicine;14(10)

https://doi.org/10.3390/jcm14103545

Citació recomanada

Aquesta citació s'ha generat automàticament.

Drets

Attribution 4.0 International

http://creativecommons.org/licenses/by/4.0/

Aquest element apareix en la col·lecció o col·leccions següent(s)