Calibration of achromatic Fresnel rhombs with an elliptical retarder model in Mueller matrix ellipsometers

Data de publicació

2025-04-11T13:30:00Z

2025-04-11T13:30:00Z

2022

2025-04-11T13:30:00Z

Resum

Fresnel rhombs are the most achromatic form of retarders available and they are suitable compensating elements for spectroscopic Mueller matrix ellipsometers. However, the small stress in the rhomb caused by the mount or produced during the fabrication can affect the ellipsometry measurements with non-negligible systematic errors, due to the relatively long path length of light inside the rhomb. This work describes a calibration method that considers the non-ideal response of the Fresnel rhomb, and that it is especially well suited for calibrating Mueller matrix ellipsometers. The method describes each rhomb as the most general form of an elliptical retarder with a small ellipticity, instead of simply assuming that they behave as linear retarders.After this calibration, we show that the systematic errors of measurements are significantly decreased.

Tipus de document

Article


Versió acceptada

Llengua

Anglès

Publicat per

Elsevier B.V.

Documents relacionats

Versió postprint del document publicat a: https://doi.org/10.1016/j.tsf.2022.139581

Thin Solid Films, 2022, vol. 763

https://doi.org/10.1016/j.tsf.2022.139581

Citació recomanada

Aquesta citació s'ha generat automàticament.

Drets

cc-by-nc-nd (c) Elsevier B.V., 2022

http://creativecommons.org/licenses/by-nc-nd/4.0/